Tracking and separation of relativistic ions using Timepix3 with a 300 um thick silicon sensor

dc.contributor.authorSmolyanskiy, P.I.
dc.contributor.authorAzzarello, Philipp
dc.contributor.authorBergmann, Benedikt
dc.contributor.authorBurian, Petr
dc.contributor.authorBroulím, Pavel
dc.contributor.authorMeduna, Lukáš
dc.contributor.authorPaniccia, Mercedes
dc.contributor.authorPerrina, Chiara
dc.contributor.authorPospíšil, Stanislav
dc.contributor.authorWu, X.
dc.date.accessioned2021-08-30T10:00:15Z
dc.date.available2021-08-30T10:00:15Z
dc.date.issued2021
dc.description.abstract-translatedThe response of a Timepix3 detector (256 x 256 pixels, pixel pitch 55 μm) with a 300 μm thick silicon sensor was studied in relativistic charged particle beams at the Super-Proton- Synchrotron at CERN. The detector was irradiated at different angles in a 400 GeV/𝑐 primary proton beam and in a mixed beam created by a 330 GeV/𝑐 Pb beam hitting a beryllium target. We present and discuss energy deposition spectra and particle track features for relativistic particles of different stopping power. The proton data shows that a relative energy resolution of approximately 9 % can be achieved. Analysis of deposited energy spectra and tracks in the detector produced by heavier ions are carried out with the aim to investigate the capabilities of the Timepix3 detector for decomposing the mixed beam. Using spectrum stripping technique by iterative Landau curve fitting charge discrimination could be done up to 𝑍 = 7 for impact angle of 60 degrees (with respect to the sensor normal). For smaller impact angles the particle charge separation is decreased due to the saturation of the pixel electronics (600 keV). Previous works describing the 3D track reconstruction for minimally ionizing particles (𝑍 = 1) were extended in the present work by adding the methodology of heavier ions track reconstruction in 3D.en
dc.format16 s.cs
dc.format.mimetypeapplication/pdf
dc.identifier.citationSMOLYANSKIY, P., AZZARELLO, P., BERGMANN, B., BURIAN, P., BROULÍM, P., MEDUNA, L., PANICCIA, M., PERRINA, CH., POSPÍŠIL, S., WU, X. Tracking and separation of relativistic ions using Timepix3 with a 300 um thick silicon sensor. Journal of Instrumentation, 2021, roč. 16, č. 1, s. 0-15. ISSN 1748-0221.cs
dc.identifier.document-number663343100017
dc.identifier.doi10.1088/1748-0221/16/01/P01022
dc.identifier.issn1748-0221
dc.identifier.obd43932001
dc.identifier.uri2-s2.0-85099776402
dc.identifier.urihttp://hdl.handle.net/11025/44953
dc.language.isoenen
dc.project.IDEF16_019/0000766/Inženýrské aplikace fyziky mikrosvětacs
dc.publisherIOP Publishingen
dc.relation.ispartofseriesJournal of Instrumentationen
dc.rightsPlný text je přístupný v rámci univerzity přihlášeným uživatelům.cs
dc.rights© IOP Publishingen
dc.rights.accessrestrictedAccessen
dc.subject.translatedhybrid detectorsen
dc.subject.translatedparticle tracking detectors (solid-state detectors)en
dc.subject.translateddE/dx detectorsen
dc.subject.translatedheavy-ion detectorsen
dc.titleTracking and separation of relativistic ions using Timepix3 with a 300 um thick silicon sensoren