Pressure Sensor Catheter based on Micromachined Aluminum Nitride Membrane Fiber Tip Fabry-Pérot-Interferometer

dc.contributor.authorFriedemann, Marvin
dc.contributor.authorMeinel, Katja
dc.contributor.authorVoigt, Sebastian
dc.contributor.authorHecker, Raoul
dc.contributor.authorKriebel, David
dc.contributor.authorHiller, Karla
dc.contributor.authorStöckel, Chris
dc.contributor.authorMehner, Jan
dc.contributor.editorPinker, Jiří
dc.date.accessioned2021-10-27T10:11:01Z
dc.date.available2021-10-27T10:11:01Z
dc.date.issued2021
dc.description.abstract-translatedThis paper introduces an absolute pressure sensor based on an extrinsic fiber-optic Fabry-Pérot-Interferometer molded into a silicone catheter. The interferometer principle is formed between the tip of an optical fiber and a reflective multilayer membrane containing aluminum nitride. The signal analysis is done by a Fiber Bragg Grating interrogation device. First, the optical sensing principle as pressure sensor in an invasive catheter with a diameter of 1 mm is described. Then, the dependence of the optical signal on changes of hydrostatic pressure and ambient temperature is shown. Out of these characteristics it was possible to evaluate the pressure sensor catheter for potential application in the human vascular system.en
dc.format4 s.cs
dc.format.mimetypeapplication/pdf
dc.identifier.citation2021 International Conference on Applied Electronics: Pilsen, 7th – 8th September 2021, Czech Republic, p. 43-46.en
dc.identifier.isbn978–80–261–0972–3 (Print)
dc.identifier.isbn978–80–261–0973–0 (Online)
dc.identifier.issn1803–7232 (Print)
dc.identifier.issn1805–9597 (Online)
dc.identifier.urihttp://hdl.handle.net/11025/45563
dc.language.isoenen
dc.publisherUniversity of West Bohemiaen
dc.rights© University of West Bohemia, 2021en
dc.rights.accessopenAccessen
dc.subjectoptické senzorycs
dc.subjectFabry–Pérot interferometrcs
dc.subjecthydrostatický tlakcs
dc.subjectteplotacs
dc.subjectvýslechové zařízenícs
dc.subjectAlN membránacs
dc.subjectkatétrcs
dc.subject.translatedfiber-optic sensorsen
dc.subject.translatedFabry-Pérot-Interferometeren
dc.subject.translatedhydrostatic pressureen
dc.subject.translatedtemperatureen
dc.subject.translatedinterrogation deviceen
dc.subject.translatedAlN membraneen
dc.subject.translatedcatheteren
dc.titlePressure Sensor Catheter based on Micromachined Aluminum Nitride Membrane Fiber Tip Fabry-Pérot-Interferometeren
dc.typeconferenceObjecten
dc.typekonferenční příspěvekcs
dc.type.statusPeer-revieweden
dc.type.versionpublishedVersionen

Files

Original bundle
Showing 1 - 1 out of 1 results
No Thumbnail Available
Name:
Pressure_Sensor_Catheter_based_on_Micromachined_Aluminum_Nitride_Membrane_Fiber_Tip_Fabry-Prot-Interferometer.pdf
Size:
967.09 KB
Format:
Adobe Portable Document Format
Description:
Plný text
License bundle
Showing 1 - 1 out of 1 results
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: