Monitoring of a skin temperature in vicinity of a surgical wound
dc.contributor.author | Tolar, David | |
dc.contributor.author | Štork, Milan | |
dc.contributor.editor | Pinker, Jiří | |
dc.date.accessioned | 2019-10-09T09:07:53Z | |
dc.date.available | 2019-10-09T09:07:53Z | |
dc.date.issued | 2016 | |
dc.description.abstract-translated | This article describes experimental measurements of a skin temperature in vicinity of a surgical wound. Measurements were carried out in collaboration with the orthopedic department of the University Hospital Olomouc, Czech Republic. A system used for sensing, transmission and visualization of physiological parameters was developed at the University of West Bohemia in Pilsen. A main goal of measurements was to evaluate technical parameters of the system, and also non-technical aspects associated with the use of the system. Technical parameters that was verified was a reliability of the system as a whole. This covers a temperature and motion sensing, a wireless transfer of measured data from sensors through wireless routers to the data server and data visualization in web application. To asses non-technical aspects of the system use, a sociological survey was carried out. Patients were asked to answer several questions related to the use of the system and temperature sensors. Conducted series of measurements and the survey showed that the vast majority of patients are accepting new technology positively. Measurements also verified the usability of the temperature sensor for local skin temperature measurement. Furthermore measurements revealed some issues associated with use of the sensor for monitoring of human body temperature and set a new target for further research and development | en |
dc.format | 4 s. | cs |
dc.format.mimetype | application/pdf | |
dc.identifier.citation | 2016 International Conference on Applied Electronics: Pilsen, 6th – 7th September 2016, Czech Republic, p.273-276. | en |
dc.identifier.isbn | 978–80–261–0601–2 (Print) | |
dc.identifier.isbn | 978–80–261–0602–9 (Online) | |
dc.identifier.issn | 1803–7232 (Print) | |
dc.identifier.issn | 1805–9597 (Online) | |
dc.identifier.uri | http://hdl.handle.net/11025/35297 | |
dc.language.iso | en | en |
dc.publisher | Západočeská univerzita v Plzni | cs |
dc.rights | © Západočeská univerzita v Plzni | cs |
dc.rights.access | openAccess | en |
dc.subject | měření teploty | cs |
dc.subject | teplotní senzory | cs |
dc.subject | kůže | cs |
dc.subject | monitoring | cs |
dc.subject | zranění | cs |
dc.subject | bezdrátová komunikace | cs |
dc.subject.translated | temperature measurement | en |
dc.subject.translated | temperature sensors | en |
dc.subject.translated | skin | en |
dc.subject.translated | monitoring | en |
dc.subject.translated | wounds | en |
dc.subject.translated | wireless communication | en |
dc.title | Monitoring of a skin temperature in vicinity of a surgical wound | en |
dc.type | konferenční příspěvek | cs |
dc.type | conferenceObject | en |
dc.type.status | Peer-reviewed | en |
dc.type.version | publishedVersion | en |
Files
Original bundle
1 - 1 out of 1 results
No Thumbnail Available
- Name:
- Tolar.pdf
- Size:
- 311.12 KB
- Format:
- Adobe Portable Document Format
- Description:
- Plný text
License bundle
1 - 1 out of 1 results
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.71 KB
- Format:
- Item-specific license agreed upon to submission
- Description: