Strain-stress analysis of lower limb with applied fixator
Files
Date issued
2010
Journal Title
Journal ISSN
Volume Title
Publisher
University of West Bohemia
Abstract
This paper compares physiological state of tibia before and after application of an external fixator. The fixator
systems’ models but also model of tibia are loaded in the direction of body axis. The paper is focused on the
examination of differences in stiffness before and after the application of fixation. Two types of axial external
fixators are compared. Both fixators differ in their construction. The first fixator is two-frame and fixation rods
are used for fixing the bone tissue (variant I). The second one is fixed into tibia with screws (variant II). We have
found out that the two-frame external fixator has much bigger stiffness during limb fixation than the fixator with
one body. Much higher deformations compared to physiological state of tibia occur in the variant II.