Influence of selected parameters on detection of heart valves outlet strut fractures usinge Ect method

dc.contributor.authorČápová, Klára
dc.contributor.authorStrapáčová, Tatiana
dc.contributor.editorKarban, Pavel
dc.date.accessioned2017-04-04T12:12:51Z
dc.date.available2017-04-04T12:12:51Z
dc.date.issued2009
dc.description.abstract-translatedThe Bjork-Shiley Convexo Concave (BSCC) mechanical heart valve has been used in surgery between 1979- 1986. It was one of the most prevalent mechanical heart valves in that period. The valve is however plagued by an increased occurrence of unexpected mechanical failures of its outlet strut compared to equivalent valve types, with a high incidence of mortality, when they occur. The periodic evaluation of the state of the BSCC valve, non-destructive test of each individual valve may help to detect the fractures presented in heart valve replacements and so prevent the damages of the valve. According to these facts this article presents one of the electromagnetic methods - eddy current testing for noninvasive inspection of BSCC valve integrity. Prior studies of eddy-current testing have shown that the detection capabilities of the method depend on various parameters. Influences of the liftoff and excitation frequency on detected signal originated from cracked outlet strut are investigated and presented. The first part of the article contains introduction of the problem and description of the defect that occurs in BSCC heart valve. Description of the simulated problem that was performed, obtained results and their interpretation in the medicine are presented in the second part of the article.en
dc.format5 s.cs
dc.format.mimetypeapplication/pdf
dc.identifier.citationAMTEE ’09 : eighth international conference on Advanced Methods in the Theory of Electrical Engineering : September 07-09, 2009 [Cheb, Czech Republic].en
dc.identifier.isbn978-80-7043-821-3
dc.identifier.urihttp://hdl.handle.net/11025/25869
dc.language.isoenen
dc.publisherUniversity of West Bohemiaen
dc.rights© University of West Bohemiaen
dc.rights.accessopenAccessen
dc.subjectBjork-Shileyho protetické srdeční chlopněcs
dc.subjectmateriál s defektní výstupní vzpěroucs
dc.subjecttestování vířivými proudycs
dc.subjectnumerická simulacecs
dc.subject.translatedBjork-Shiley prosthetic heart valvesen
dc.subject.translatedmaterial defect of outlet struten
dc.subject.translatededdy current testingen
dc.subject.translatednumerical simulationen
dc.titleInfluence of selected parameters on detection of heart valves outlet strut fractures usinge Ect methoden
dc.typekonferenční příspěvekcs
dc.typeconferenceObjecten
dc.type.statusPeer-revieweden
dc.type.versionpublishedVersionen

Files

Original bundle
Showing 1 - 1 out of 1 results
No Thumbnail Available
Name:
Strapacova.pdf
Size:
280.54 KB
Format:
Adobe Portable Document Format
Description:
Plný text
License bundle
Showing 1 - 1 out of 1 results
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: